Hemolysis Index and Potassium Reporting

Am J Clin Pathol. 2022 Jun 7;157(6):809-813. doi: 10.1093/ajcp/aqab217.

Abstract

Objectives: In vitro hemolysis generates a spurious increase in potassium. Roche Diagnostics recently revised its recommended guidelines for potassium reporting on cobas analyzers. By dramatically reducing the allowable degree of hemolysis, these guidelines would increase specimen rejection rates. We attempted to balance the desire to avoid inaccurate results with the clinical implications of increased specimen rejection rates.

Methods: We downloaded hemolytic indices (HI) for 80,795 specimens tested at our institution on cobas chemistry analyzers in 1 month and evaluated potential specimen rejection rates based on the new criteria. We also spiked nonhemolyzed samples with hemolyzed blood to assess the influence of HI values on potassium measurements.

Results: The new recommendations would lead to specimen rejection rates of 76% in the neonatal intensive care unit (NICU), 41% in the emergency department (ED), 16% in inpatient specimens, and 9% in outpatient samples. Our current criteria of reporting potassium concentrations in inpatient and outpatient specimens with HI ≤100 and in NICU and ED specimens with HI ≤300 and additional interpretive guidance for HI values between 100 and 300 reduce unnecessary specimen rejections to 3% in NICU, 2% in ED and inpatients, and less than 1% in outpatients without significantly increasing the number of clinically consequential incorrect results.

Conclusions: The new recommendations would lead to unacceptably high specimen rejection rates. Laboratories should develop context-specific, evidence-based reporting criteria that minimize reporting of inaccurate results without disrupting delivery of care.

Keywords: Guidelines; Hemolytic index; Potassium; Reporting.

MeSH terms

  • Hematologic Tests
  • Hemolysis*
  • Humans
  • Infant, Newborn
  • Laboratories
  • Potassium*

Substances

  • Potassium