Highly Sensitive Protein Detection by Asymmetric Mach-Zehnder Interferometry for Biosensing Applications

ACS Appl Bio Mater. 2020 Jul 20;3(7):4566-4572. doi: 10.1021/acsabm.0c00491. Epub 2020 Jun 25.

Abstract

The sensitivity and performance of an asymmetric Mach-Zehnder interferometer (aMZI) were compared to those of quartz crystal microbalance with dissipation (QCM-D). The binding of streptavidin to sensor chips coated with poly-l-lysine (PLL), modified with biotin and oligoethyleneglycol (OEG) (PLL-biotin), was used to compare the binding signals obtained from both technologies. PLL-biotin proved to be an efficient method to add bioreceptors to both the QCM-D and aMZI chips. The final, saturated value of streptavidin binding was compared with those from aMZI (253 ng cm-2) and QCM-D (460 ng cm-2). These values were then used to evaluate that 45% of the measured streptavidin mass in the QCM-D came from hydrodynamically coupled water. Importantly, the signal-to-noise ratio of the aMZI was found to be 200 times higher than that of the QCM-D. These results indicate the potential of the aMZI platform for highly sensitive and accurate biosensing applications.

Keywords: asymmetric Mach−Zehnder interferometry; biosensing; poly-l-lysine; quartz crystal microbalance; surface plasmon resonance.