Robot-guided versus freehand fluoroscopy-guided minimally invasive transforaminal lumbar interbody fusion: a single-institution, observational, case-control study

Neurosurg Focus. 2022 Jan;52(1):E9. doi: 10.3171/2021.10.FOCUS21514.

Abstract

Objective: The use of robotics in spinal surgery has gained popularity because of its promising accuracy and safety. ROSA is a commonly used surgical robot system for spinal surgery. The aim of this study was to compare outcomes between robot-guided and freehand fluoroscopy-guided instrumentation in minimally invasive surgery (MIS)-transforaminal lumbar interbody fusion (TLIF).

Methods: This retrospective consecutive series reviewed 224 patients who underwent MIS-TLIF from March 2019 to April 2020 at a single institution. All patients were diagnosed with degenerative pathologies. Of those, 75 patients underwent robot-guided MIS-TLIF, and 149 patients underwent freehand fluoroscopy-guided MIS-TLIF. The incidences of pedicle breach, intraoperative outcomes, postoperative outcomes, and short-term pain control were compared.

Results: The patients who underwent robot-guided surgery had a lower incidence of pedicle breach (0.27% vs 1.75%, p = 0.04) and less operative blood loss (313.7 ± 214.1 mL vs 431.6 ± 529.8 mL, p = 0.019). Nonsignificant differences were observed in operative duration (280.7 ± 98.1 minutes vs 251.4 ± 112.0 minutes, p = 0.056), hospital stay (6.6 ± 3.4 days vs 7.3 ± 4.4 days, p = 0.19), complications (intraoperative, 1.3% vs 1.3%, p = 0.45; postoperative surgery-related, 4.0% vs 4.0%, p = 0.99), and short-term pain control (postoperative day 1, 2.1 ± 1.2 vs 1.8 ± 1.2, p = 0.144; postoperative day 30, 1.2 ± 0.5 vs 1.3 ± 0.7, p = 0.610). A shorter operative duration for 4-level spinal surgery was found in the robot-guided surgery group (388.7 ± 107.3 minutes vs 544.0 ± 128.5 minutes, p = 0.047).

Conclusions: This retrospective review revealed that patients who underwent robot-guided MIS-TLIF experienced less operative blood loss. They also benefited from a shorter operative duration with higher-level (> 3 levels) spinal surgery. The postoperative outcomes were similar for both robot-guided and freehand fluoroscopy-guided procedures.

Keywords: TLI; pedicle scre; robo; transforaminal lumbar interbody fusio.

Publication types

  • Observational Study

MeSH terms

  • Case-Control Studies
  • Fluoroscopy
  • Humans
  • Lumbar Vertebrae / diagnostic imaging
  • Lumbar Vertebrae / surgery
  • Minimally Invasive Surgical Procedures / methods
  • Retrospective Studies
  • Robotics*
  • Spinal Fusion* / methods
  • Treatment Outcome