Pocket Mercury-Vapour Detection System Employing a Preconcentrator Based on Au-TiO2 Nanomaterials

Sensors (Basel). 2021 Dec 10;21(24):8255. doi: 10.3390/s21248255.

Abstract

In environments polluted by mercury vapors that are potentially harmful to human health, there is a need to perform rapid surveys in order to promptly identify the sources of emission. With this aim, in this work, a low cost, pocket-sized portable mercury measurement system, with a fast response signal is presented. It consists of a preconcentrator, able to adsorb and subsequently release the mercury vapour detected by a quartz crystal microbalance (QCM) sensor. The preconcentrator is based on an adsorbing layer of titania/gold nanoparticles (TiO2NP/AuNPs), deposited on a micro-heater that acts as mercury thermal desorption. For the detection of the released mercury vapour, gold electrodes QCM (20 MHz) have been used. The experimental results, performed in simulated polluted mercury-vapour environments, showed a detection capability with a prompt response. In particular, frequency shifts (-118 Hz ± 2 Hz and -30 Hz ± 2 Hz) were detected at concentrations of 65 µg/m3 Hg0 and 30 µg/m3 Hg0, with sampling times of 60 min and 30 min, respectively. A system limit of detection (LOD) of 5 µg/m3 was evaluated for the 30 min sampling time.

Keywords: Arduino; QCM; low cost; mercury; portable system; sensors.

MeSH terms

  • Gold
  • Humans
  • Mercury*
  • Metal Nanoparticles*
  • Quartz Crystal Microbalance Techniques
  • Titanium

Substances

  • titanium dioxide
  • Gold
  • Titanium
  • Mercury