Computed Tomography-Guided Placement of Sacral Electrodes

Isr Med Assoc J. 2021 Dec;23(12):773-776.

Abstract

Background: The current approach to performing sacral neuromodulation consists of a two-stage procedure, the first of which includes insertion of the sacral electrode under fluoroscopic visualization of the S3 foramen. Alternatively, in certain situations computed tomography (CT)-guided insertion can be used.

Objectives: To evaluate the use of CT in cases of reinsertion of the electrode due to infection, dislocation, or rupture.

Methods: Medical records of patients who underwent neuromodulation device reinsertion between 2005 and 2016 for fecal incontinence were reviewed. Study outcomes included procedure course, successful placement, and long-term treatment success.

Results: During the study period, we inserted a neuromodulation device in 67 patients. A CT-guided insertion of a sacral electrode was performed in 10 patients. In nine patients, the insertion and the final location of the electrode were successful. In one patient, the electrode migrated upward due to a malformation of the S3 foramen on both sides and had to be placed in S4. In a mean follow-up of 68.4 ± 30.0 months following the re-insertion, there was a significant reduction in the number of incontinence episodes per day (P < 0.001) and the number of pads used per day (P = 0.002).

Conclusions: CT-guided insertion of a sacral electrode is a safe and promising option, especially in recurrent and or selected cases.

MeSH terms

  • Adolescent
  • Aged
  • Electric Stimulation Therapy / methods*
  • Electrodes, Implanted
  • Fecal Incontinence / therapy*
  • Female
  • Follow-Up Studies
  • Humans
  • Male
  • Middle Aged
  • Sacrum / diagnostic imaging*
  • Sacrum / innervation
  • Tomography, X-Ray Computed / methods*