Development of a humidity pretreatment method for the measurement of ozone in ambient air

J Hazard Mater. 2022 Mar 15:426:128108. doi: 10.1016/j.jhazmat.2021.128108. Epub 2021 Dec 18.

Abstract

A frost filter (FRF) was developed as a humidity pretreatment device (HPD) to improve the measurement of ambient ozone (O3). The FRF was produced in a tube, which was supercooled by a thermoelectric cooling device based on the Peltier effect. The relative humidity (RH) of the air samples varied from 30% to 80% at 25 °C, and the O3 concentration was set as 100 ppbv. Besides O3, SO2 at 150 ppbv was used for comparison. The density of the FRF was evaluated. Comparison studies on the humidity removal efficiencies and loss ratios of analytes among a FRF HPD, a short Nafion™ tube (NS), and a long Nafion™ tube (NL) HPDs were conducted. As results, the density of the FRF was dependent on the temperature at a fixed sampling flow rate. The outlet humidity of both the FRF and the NL HPDs were less than 8% RH at 25 °C. The mean concentrations of O3 and SO2 after the FRF HPD were similar to the initial concentrations at all humidity levels, whereas they were significantly different for both the NS and NL HPDs at higher humidity. This suggests that the FRF HPD is a reliable humidity pretreatment for O3 measurements.

Keywords: Dehumidification; Frost filter; Monitoring; Nafion; Permeation.