Reliability of resting metabolic rate between and within day measurements using the Vyntus CPX system and comparison against predictive formulas

Nutr Health. 2023 Mar;29(1):107-114. doi: 10.1177/02601060211057324. Epub 2021 Dec 21.

Abstract

Background: To detect longitudinal changes of resting metabolic rate (RMR) resulting from the effects of energetic stress, reliable RMR measurements are crucial. The Vyntus CPX is a new automated indirect calorimetry system for which RMR reliability has not been determined. Additionally, its agreement with common predictive RMR formulas is unknown. Aim: To determine the within and between-day reliability of RMR measurements using the Vyntus CPX system and its agreement with predictive RMR formulas. Methods: Young (31 ± 7 years) healthy participants (n = 26, 12 females, 14 males) completed three measurements of RMR, two consecutive measures on the same day, one the day before/after, all under standardised conditions. Reliability was assessed with pairwise comparisons of between-day at the same time (BDST), within day consecutive measurements (WDCM) and between-day different time (BDDT), for parameters of reliability (mean change (MC), intraclass correlation (ICC) and typical error of measurement (TEM)). Measured RMR values (kcal/day) were compared against predictive values of 4 common formulas. Results: Parameters of reliability (mean, (95% confidence interval)) were: -BDST: MC, 0.2(-2.3-2.7)% (p = 0.67); ICC, 0.92(0.84-0.97); TEM, 4.5(3.5-6.2)%. -WDCM: MC, -2.5(-6.2-1.3)% (p = 0.21); ICC, 0.88(0.74-0.88); TEM, 7.0(5.4-9.8)%. -BDDT: MC, -1.5(-4.8-1.9)% (p = 0.57); ICC, 0.90(0.76-0.95); TEM, 6.1(4.8-8.5)%. RMRratios (measured/predicted) were: 1.04 ± 0.14 (Nelson, p = 0.13), 1.03 ± 0.10 (Mifflin, p = 0.21), 0.98 ± 0.09 (Harris-benedict, p = 0.30), 0.95 ± 0.11 (Cunningham1980, p = 0.01), 1.00 ± 0.12 (Cunningham1991, p = 0.90) and 0.96 ± 0.13 (DXA, p = 0.03). Conclusions: The Vyntus CPX is reliable and measured RMR values agreed with four predictive formulas but are lower than Cunningham1980 and DXA RMR estimates for this population.

Keywords: RMRratio; indirect calorimetry; reliability; resting metabolic rate.

MeSH terms

  • Basal Metabolism*
  • Calorimetry, Indirect / methods
  • Female
  • Healthy Volunteers
  • Humans
  • Male
  • Reproducibility of Results