Personalised reprogramming to prevent progressive pacemaker-related left ventricular dysfunction: A phase II randomised, controlled clinical trial

PLoS One. 2021 Dec 13;16(12):e0259450. doi: 10.1371/journal.pone.0259450. eCollection 2021.

Abstract

Background: Pacemakers are widely utilised to treat bradycardia, but right ventricular (RV) pacing is associated with heightened risk of left ventricular (LV) systolic dysfunction and heart failure. We aimed to compare personalised pacemaker reprogramming to avoid RV pacing with usual care on echocardiographic and patient-orientated outcomes.

Methods: A prospective phase II randomised, double-blind, parallel-group trial in 100 patients with a pacemaker implanted for indications other than third degree heart block for ≥2 years. Personalised pacemaker reprogramming was guided by a published protocol. Primary outcome was change in LV ejection fraction on echocardiography after 6 months. Secondary outcomes included LV remodeling, quality of life, and battery longevity.

Results: Clinical and pacemaker variables were similar between groups. The mean age (SD) of participants was 76 (+/-9) years and 71% were male. Nine patients withdrew due to concurrent illness, leaving 91 patients in the intention-to-treat analysis. At 6 months, personalised programming compared to usual care, reduced RV pacing (-6.5±1.8% versus -0.21±1.7%; p<0.01), improved LV function (LV ejection fraction +3.09% [95% confidence interval (CI) 0.48 to 5.70%; p = 0.02]) and LV dimensions (LV end systolic volume indexed to body surface area -2.99mL/m2 [95% CI -5.69 to -0.29; p = 0.03]). Intervention also preserved battery longevity by approximately 5 months (+0.38 years [95% CI 0.14 to 0.62; p<0.01)) with no evidence of an effect on quality of life (+0.19, [95% CI -0.25 to 0.62; p = 0.402]).

Conclusions: Personalised programming in patients with pacemakers for bradycardia can improve LV function and size, extend battery longevity, and is safe and acceptable to patients.

Trial registration: ClinicalTrials.gov identifier: NCT03627585.

Publication types

  • Clinical Trial, Phase II
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Bradycardia / therapy
  • Double-Blind Method
  • Echocardiography
  • Female
  • Heart Ventricles / physiopathology
  • Humans
  • Male
  • Natriuretic Peptide, Brain / blood
  • Outcome Assessment, Health Care
  • Pacemaker, Artificial / adverse effects*
  • Patient-Specific Modeling
  • Peptide Fragments / blood
  • Quality of Life
  • Stroke Volume
  • Ventricular Dysfunction, Left / etiology
  • Ventricular Dysfunction, Left / prevention & control*
  • Ventricular Remodeling*

Substances

  • Peptide Fragments
  • pro-brain natriuretic peptide (1-76)
  • Natriuretic Peptide, Brain

Associated data

  • ClinicalTrials.gov/NCT03627585