A highly accurate mass spectrometry method for the quantification of phenylalanine and tyrosine on dried blood spots: Combination of liquid chromatography, phenylalanine/tyrosine-free blood calibrators and multi-point/dynamic calibration

Clin Biochem. 2022 Mar:101:35-41. doi: 10.1016/j.clinbiochem.2021.12.004. Epub 2021 Dec 8.

Abstract

Background: Measurement of quantitative levels of phenylalanine and tyrosine in blood is an essential test for the diagnosis of and monitoring genetic disorders associated with phenylalanine metabolism, such as phenylketonuria (PKU), tyrosinemia, and defects of tetrahydrobiopterin synthesis and recycling. We developed a highly accurate and fast liquid chromatography with tandem mass spectrometry (LC-MS/MS) method for the quantification of phenylalanine and tyrosine on dried blood spot (DBS). We also designed a performance score system to evaluate various calibration methods in matrix matched material.

Methods: Phenylalanine/tyrosine-free whole blood was used to make accurate and stable DBS calibrators. Six calibrators cover the range of 0-1000 µmol/L. Underivatized phenylalanine and tyrosine were extracted and measured by LC-MS/MS. Precision, accuracy, limit of quantification, recovery and carryover were validated. External quality assurance materials were also used to evaluate performance of multi-point calibrations and single-point calibrations.

Results: The run time was 4.5-minute. Accuracy analysis showed good agreement with reference materials. Precision, recovery, and the lower and upper limit of quantification met the criteria. When phenylalanine and tyrosine concentrations were less than 150 µmol/L, the 5-point calibration without the calibrator of 1000 µmol/L had the best performance. When the concentrations were > 250 µmol/L, the single-point calibration of 500 µmol/L had the best performance.

Conclusion: We developed a simple, fast and highly accurate method for the detection of phenylalanine and tyrosine on DBS, with chromatographic separation and underivatized analysis. Based on the calibration performance, a 6-point calibration method is satisfying for this test. An optional dynamic calibration method, which includes 6-point calibration, 5-point calibration and single-point calibration, can further increase test reliability.

Keywords: LC-MS/MS; Multi-point calibration; Phenylalanine; Tyrosine; Underivatization.