Multi-Detector Heart Rate Extraction Method for Transabdominal Fetal Pulse Oximetry

Annu Int Conf IEEE Eng Med Biol Soc. 2021 Nov:2021:1072-1075. doi: 10.1109/EMBC46164.2021.9630946.

Abstract

Intrapartum fetal well-being assessment relies on fetal heart rate (FHR) monitoring. Studies have shown that FHR monitoring has a high false-positive rate for detecting fetal hypoxia during labor and delivery. A transabdominal fetal pulse oximeter device that measures fetal oxygen saturation non-invasively through NIR light source and photodetectors could increase the accuracy of hypoxia detection. As light travels through both maternal and fetal tissue, photodetectors on the surface of mother's abdomen capture mixed signals comprising fetal and maternal information. The fetal information should be extracted first to enable fetal oxygen saturation calculation. A multi-detector fetal signal extraction method is presented in this paper where adaptive noise cancellation is applied to four mixed signals captured by four separate photodetectors placed at varying distances from the light source. As a result of adaptive noise cancellation, we obtain four separate FHR by peak detection. Weighting, outlier rejection and averaging are applied to these four fetal heart rates and a mean FHR is reported. The method is evaluated in utero on data collected from hypoxic lamb model. Ground truth for FHR is measured through hemodynamics. The results showed that using multi-detector fetal signal extraction gave up to 18.56% lower root-mean-square FHR error, and up to 57.87% lower maximum absolute FHR error compared to single-detector fetal signal extraction.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Animals
  • Female
  • Fetal Hypoxia
  • Heart Rate, Fetal*
  • Oximetry*
  • Oxygen
  • Oxygen Saturation
  • Pregnancy
  • Sheep

Substances

  • Oxygen