Results from a multicentre evaluation of plug use for left ventricular assist device explantation

Interact Cardiovasc Thorac Surg. 2022 Mar 31;34(4):683-690. doi: 10.1093/icvts/ivab344.

Abstract

Objectives: Myocardial recovery allows for left ventricular assist device (LVAD) explantations after long-term support. Several surgical approaches, including interventional decommissioning, off-pump explantation using a custom-made plug and complete LVAD removal through redo sternotomy, have been described. We present the results from an evaluation of the long-term follow-up of patients who received a titanium sintered plug after LVAD explantation.

Methods: We performed a retrospective, European, multicentre analysis of patients who received a titanium sintered plug to seal the apical fixation ring after LVAD explantation. Data were collected from a questionnaire that included demographics, procedural details and follow-up information.

Results: Out of 54 contacted centres in 12 countries (n = 179 patients), a total of 68 patients were successfully included in the study. The median follow-up was 34 months (interquartile range: 17-58.5 months); 57 (84%) patients had >1-year follow-up. At the time of the last follow-up, 55 (81%) patients were alive, with a Kaplan-Meier 1-year survival of 90.1% (95% confidence interval: 84.0-98.1%) and a 5-year survival of 80.0% (95% confidence interval: 68.4-92.9%). One patient (1.5%) developed a plug infection originating from an infected part of the incorporated driveline and, after complete removal, is currently in good condition. No postoperative stroke has been reported after plug implantation.

Conclusions: In this European multicentre study, the use of a custom-made titanium plug to close the apical fixation ring after LVAD explantation resulted in a low incidence of plug-related complications. With the volume of patients undergoing LVAD explantations after myocardial recovery increasing, the plug has evolved as a simple alternative to more invasive device explantation procedures or decommissioning with a high risk for infection of the remaining system or stroke.

Keywords: Explantation; Left ventricular assist device; MCS; Plug; Recovery; Weaning.

Publication types

  • Multicenter Study

MeSH terms

  • Device Removal / adverse effects
  • Device Removal / methods
  • Heart Failure* / surgery
  • Heart-Assist Devices* / adverse effects
  • Humans
  • Retrospective Studies
  • Treatment Outcome