The artificial iris - Analysis of various implantation techniques after ocular trauma

Indian J Ophthalmol. 2021 Dec;69(12):3526-3531. doi: 10.4103/ijo.IJO_62_21.

Abstract

Purpose: The aim of this study is to analyze the outcome of various techniques for a custom-made iris prosthesis implantation as part of reconstructive anterior segment surgery following traumatic aniridia.

Methods: This retrospective interventional study was done for 6 eyes that received an artificial iris as secondary reconstructive measure for photophobia and unsatisfactory vision following initial globe repair. Different implantation techniques were employed. These included simple sulcus implantation, implantation of a composite (iris prosthesis with attached intraocular lens) implant, and combinations with phacoemulsification, vitrectomy, and penetrating keratoplasty.

Results: In all cases, the artificial iris was implanted successfully. In the follow-up period (1-48 months), postoperative complications included rhegmatogenous retinal detachment, prolonged intraocular inflammation, and corneal transplant decompensation due to graft rejection. There was no case of secondary glaucoma. Complications could be managed successfully. All patients showed improved best-corrected visual acuity and were satisfied with functional and cosmetic results.

Conclusion: This case series highlights the different implantation techniques for reconstruction of the anterior segment after ocular trauma. The versatility of the custom-made iris implant accounts for a wide range of applications and the foldable material reduces the need for large incisions in the already traumatized eye.

Keywords: Artificial Iris; iris prosthesis; secondary intraocular lens; traumatic aniridia.

MeSH terms

  • Aniridia* / surgery
  • Eye Injuries* / complications
  • Eye Injuries* / diagnosis
  • Eye Injuries* / surgery
  • Humans
  • Iris / surgery
  • Lens Implantation, Intraocular
  • Lenses, Intraocular*
  • Retrospective Studies
  • Visual Acuity