Motion in nuclear cardiology imaging: types, artifacts, detection and correction techniques

Phys Med Biol. 2022 Jan 17;67(2). doi: 10.1088/1361-6560/ac3dc7.

Abstract

In this paper, the authors review the field of motion detection and correction in nuclear cardiology with single photon emission computed tomography (SPECT) and positron emission tomography (PET) imaging systems. We start with a brief overview of nuclear cardiology applications and description of SPECT and PET imaging systems, then explaining the different types of motion and their related artefacts. Moreover, we classify and describe various techniques for motion detection and correction, discussing their potential advantages including reference to metrics and tasks, particularly towards improvements in image quality and diagnostic performance. In addition, we emphasize limitations encountered in different motion detection and correction methods that may challenge routine clinical applications and diagnostic performance.

Keywords: PET; SPECT; motion correction; motion detection; nuclear cardiology; patient motion.

Publication types

  • Research Support, Non-U.S. Gov't
  • Review

MeSH terms

  • Artifacts*
  • Cardiology*
  • Motion
  • Positron-Emission Tomography / methods
  • Tomography, X-Ray Computed / methods