Ultrasonic Evaluation of the Heel Fat Pad under Weight-Bearing Conditions Using a Polymethylpentene Resin Plate: Part 1

Ultrasound Med Biol. 2022 Feb;48(2):358-372. doi: 10.1016/j.ultrasmedbio.2021.10.012. Epub 2021 Nov 22.

Abstract

To evaluate the two-layer structure of the heel fat pad (HFP) from non-weight-bearing to full-weight-bearing conditions, we developed an instrument that assesses these changes from the sole through a polymethylpentene resin plate (PMP) with ultrasonography. For actual use, we investigated the influence on measured values and ultrasonogram appearance by interposing the PMP. Additionally, as the PMP may be bent under weight-bearing conditions, its influence on the measured values was investigated. First, two examiners measured the distances inside the phantom with and without a PMP. Second, ultrasonograms were obtained from 40 healthy adults with and without a PMP, and the thicknesses of the whole layer and the two layers of the HFP were measured using the same ultrasonogram. For each experiment, reproducibility was investigated. Third, the distances inside the phantom were measured and compared through the bent PMP, which models the weight-bearing condition, and the flat PMP. The reproducibility of the measurements was equivalent with and without the PMP interposed. Potential bias in measured values arising from deformation of PMP under weight-bearing conditions was not detected. Overall, the PMP's interposition and the bending of the PMP might not influence the measured values and reproducibility of the measurements.

Keywords: Acoustic lens; Evaluation instrument; Full-weight-bearing; Heel fat pad; Macrochamber layer; Microchamber layer; Polymethylpentene resin; Reproducibility.

MeSH terms

  • Adipose Tissue
  • Adult
  • Heel* / diagnostic imaging
  • Humans
  • Reproducibility of Results
  • Ultrasonics*
  • Weight-Bearing