High-precision 3D drift correction with differential phase contrast images

Opt Express. 2021 Oct 11;29(21):34641-34655. doi: 10.1364/OE.438160.

Abstract

Single molecule localization microscopy (SMLM) usually requires long image acquisition time at the order of minutes and thus suffers from sample drift, which deteriorates image quality. A drift estimation method with high precision is typically used in SMLM, which can be further combined with a drift compensation device to enable active microscope stabilization. Among all the reported methods, the drift estimation method based on bright-field image correlation requires no extra sample preparation or complicated modification to the imaging setup. However, the performance of this method is limited by the contrast of bright-field images, especially for the structures without sufficient features. In this paper, we proposed to use differential phase contrast (DPC) microscopy to enhance the image contrast and presented a 3D drift correction method with higher precision and robustness. This DPC-based drift correction method is suitable even for biological samples without clear morphological features. We demonstrated that this method can achieve a correction precision of < 6 nm in both the lateral direction and axial direction. Using SMLM imaging of microtubules, we verified that this method provides a comparable drift estimation performance as redundant cross-correlation.