Wear rate and osteolysis in two types of second-generation annealed highly cross-linked polyethylene in total hip arthroplasty: A retrospective comparative study with a minimum of five years

Orthop Traumatol Surg Res. 2023 Feb;109(1):103147. doi: 10.1016/j.otsr.2021.103147. Epub 2021 Nov 15.

Abstract

Background: As no previous study has directly compared the linear wear rate in two types of second-generation annealed highly cross-linked polyethylene, we performed a retrospective study with a minimum of 5-year follow-up to assess primary arthroplasties in the (1) wear rates and (2) incidence of osteolysis of the two types of HXLPE.

Hypothesis: There was no significant difference in the linear wear rate and the incidence of osteolysis between the two types of second-generation annealed highly cross-linked polyethylene.

Patients and methods: In this single-center study, we reviewed 257 cases of primary cementless total hip arthroplasties between 2011 and 2015, which were performed with 32mm delta ceramic on second-generation annealed highly cross-linked polyethylene (X3 and E1 were used in 105 and 103 cases, respectively.). The mean wear rate was evaluated using a computer-assisted method, and the incidence of osteolysis was evaluated based on the appearance of a localized area with loss of trabecular bone or cortical erosion adjacent to the implants during the latest follow-up.

Results: In total, we evaluated 208 cases, followed postoperatively for over 5 years (mean, 6.1 years, range: 5.0-8.0). There were no significant differences between the two groups with respect to age (list in order of Group X, Group E, p value) (61.2±12.3, 62.7±12.1, p=0.36), sex (ratio of male: 17.1%, 14.6%, p=0.61), body mass index (22.9±3.7, 22.8±4.0, p=0.91), pre- (49.9±14.8, 48.5±13.8, p=0.49) and post-operative (91.3±9.1, 92.7±7.0, p=0.23) Japanese Orthopaedic Association Hip Score, cup size (50.8±3.0, 50.9±2.2, p=0.70), cup inclination (38.7±4.8, 37.6±4.8, p=0.10), and cup anteversion (18.7±6.9, 18.5±7.6, p=0.80). The mean linear wear rates of the X3 and E1 groups were 0.057±0.039 (range: 0-0.16) and 0.054±0.037mm/year (range: 0-0.15), respectively (p=0.61). No osteolysis was found on the final plain radiographs in both groups.

Discussion: This study revealed that both types of highly cross-linked polyethylene have excellent linear wear rates and were equally safe to use. However, the difference between the two materials in terms of the long-term wear rate should be further validated.

Level of evidence: III; retrospective case control study.

Keywords: Cementless; Cross-linked polyethylene; E1; PolyWare; Tapered-wedge; Wear rate; X3.

Publication types

  • Review

MeSH terms

  • Arthroplasty, Replacement, Hip* / adverse effects
  • Arthroplasty, Replacement, Hip* / methods
  • Case-Control Studies
  • Follow-Up Studies
  • Hip Prosthesis* / adverse effects
  • Humans
  • Male
  • Osteolysis* / diagnostic imaging
  • Osteolysis* / epidemiology
  • Osteolysis* / etiology
  • Polyethylene
  • Prosthesis Design
  • Prosthesis Failure
  • Retrospective Studies

Substances

  • Polyethylene