Comparison of iridocorneal angle parameters measured by spectral domain optical coherence tomography and ultrasound biomicroscopy in dogs

Vet Ophthalmol. 2022 May:25 Suppl 1:103-110. doi: 10.1111/vop.12950. Epub 2021 Nov 16.

Abstract

Objective: To compare the measurements of iridocorneal angle parameters between spectral domain optical coherence tomography (SD-OCT) and ultrasound biomicroscopy (UBM) in dogs.

Animals studied and procedures: A total of 47 eyes of dogs were scanned at the temporal limbus using SD-OCT and UBM. Iridocorneal angle (ICA) and angle opening distance (AOD) were measured from the obtained images accordingly. The intra-observer and inter-observer reproducibility were evaluated using the intraclass correlation coefficient. To evaluate intra-observer reproducibility, measurements of the first and second grading from the first examiner were compared. To evaluate inter-observer reproducibility, measurements between the two examiners were compared. Agreement between ICA and AOD for SD-OCT and UBM was evaluated using Bland-Altman plots.

Results: In the first grading, the mean ICA and AOD for SD-OCT were 31.4 ± 6.4° and 641.4 ± 270.8 µm, respectively. The mean ICA and AOD for UBM were 32.0 ± 4.8° and 700.4 ± 238.8 µm, respectively. For ICA and AOD measurements, intra-observer reproducibility was excellent for both devices, whereas inter-observer reproducibility was excellent for SD-OCT and good for UBM. The mean difference in ICA between SD-OCT and UBM was 0.6° with a limit of agreement (LoA) span of 18.9°. The mean difference in AOD between SD-OCT and UBM was 58.9 µm with a LoA span of 804.4 µm.

Conclusions: Spectral domain optical coherence tomography is an effective non-contact imaging modality for the evaluation of canine iridocorneal angle parameters in a clinical setting. Reproducibility of measurements obtained is comparable or superior to UBM, but values obtained by SD-OCT and UBM for AOD are not interchangeable between devices.

Keywords: angle opening distance; dogs; iridocorneal angle; iridocorneal angle parameters; spectral domain optical coherence tomography; ultrasound biomicroscopy.