SARS-CoV-2-Impedimetric Biosensor: Virus-Imprinted Chips for Early and Rapid Diagnosis

ACS Sens. 2021 Nov 26;6(11):4098-4107. doi: 10.1021/acssensors.1c01614. Epub 2021 Nov 10.

Abstract

Due to the current global SARS-CoV-2 pandemic, rapid and accurate diagnostic tools are needed to prevent the spread of COVID-19 across the globe. An electrochemical sensing platform was constructed using CNTs/WO3-screen printed electrodes for imprinting the complete virus particles (SARS-CoV-2 particles) within the polymeric matrix to create viral complementary binding sites. The sensor provided high selectivity toward the target virus over other tested human corona and influenza respiratory interference viruses. The sensitivity performance of the sensor chips was evaluated using different viral concentrations, while the limits of detection and quantification were 57 and 175 pg/mL, respectively. Reaching this satisfied low detection limit (almost 27-fold more sensitive than the RT-PCR), the sensor was applied in clinical specimens obtained from SARS-CoV-2 suspected cases. Thus, dealing directly with clinical samples on the chip could be provided as a portable device for instantaneous and simple point of care in hospitals, airports, and hotspots.

Keywords: COVID-19; SARS-CoV-2 biosensors; carbon nanotubes/tungsten oxide; impedimetric biosensors; personal diagnostics; point of care (POC).