X-ray high frequency pulse emission characteristic and application of CNT cold cathode x-ray source cathode x-ray source

Nanotechnology. 2021 Nov 24;33(7). doi: 10.1088/1361-6528/ac378b.

Abstract

Carbon nanotube (CNT) field-emission x-ray source has great potential in x-ray communication (XCOM) because of its controllable emission and instantaneous response. A novel voltage loading mode was proposed in this work to achieve high-frequency pulse x-ray emission. The characteristics of cathode current and pulse x-ray versus voltage, frequency, and pulse amplitude were studied, and XCOM data transmission experiment was carried out. Results showed that the CNT cold cathode x-ray source, as a communication signal source, could work in 1.05 MHz pulse emission frequency. When the grid voltage was higher than 470 V, the pulse x-ray waveform amplitude achieved peak, and the shape exhibited a pseudo square wave. The duty cycle of the x-ray waveform exceeded 50%, reaching 56% when the pulse frequency reached 1 MHz. In the XCOM data transmission experiment, the pulsed x-ray waveform was well consistent with the loading data signal voltage waveform under different pulse-emission frequencies. This work realized the x-ray high-frequency pulse emission of CNT cold cathode x-ray source and lays a foundation for the development and application of CNT cold cathode x-ray source in XCOM.

Keywords: carbon nanotube x-ray source; high-frequency electron emission; pulse source; x-ray communication.