Safety-Optimized Inductive Powering of Implantable Medical Devices: Tutorial and Comprehensive Design Guide

IEEE Trans Biomed Circuits Syst. 2021 Dec;15(6):1354-1367. doi: 10.1109/TBCAS.2021.3125618. Epub 2022 Feb 17.

Abstract

A tutorial and comprehensive guide are presented for the design of planar spiral inductors with maximum energy delivery in biomedical implants. Rather than maximizing power transfer efficiency (PTE), the ratio of the received power to the square of the magnetic flux density is maximized in this technique. This ensures that the highest power is delivered for a given level of safe electromagnetic radiation, as measured by the specific absorption rate (SAR) in the tissue. By using quasi-static field approximations, the maximum deliverable power under SAR constraints is embedded in a lumped-element model of a 2-coil inductive link, from which planar coil geometries are derived. To compare the proposed methodology with the conventional approach that maximizes PTE, the results of both techniques are compared for three examples of state-of-the-art designs. It is demonstrated that the presented technique increases the maximum deliverable power while operating at a given level of non-ionizing radiation by factors of 8×, 410×, and 560× as compared to the three existing designs, and maintaining moderate link efficiencies of 12%, 23%, and 12%, respectively.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electric Power Supplies*
  • Electromagnetic Radiation
  • Equipment Design
  • Prostheses and Implants
  • Wireless Technology*

Grants and funding