Ultrasmall compact CMOS imaging system for bioluminescence reporter-based live gene expression analysis

J Biomed Opt. 2021 Nov;26(11):116002. doi: 10.1117/1.JBO.26.11.116002.

Abstract

Significance: Gene expression analysis is an important fundamental area of biomedical research. However, live gene expression imaging has proven challenging due to constraints in conventional optical devices and fluorescent reporters.

Aim: Our aim is to develop smaller, more cost-effective, and versatile imaging capabilities compared with conventional devices. Bioluminescence reporter-based gene expression analysis was targeted due to its advantages over fluorescence-based imaging.

Approach: We created a small compact imaging system using micro-CMOS image sensors (μCIS). The μCIS model had an improved pixel design and a patterned absorption filter array to detect the low light intensity of bioluminescence.

Results: The device demonstrated lower dark current, lower temporal noise, and higher sensitivity compared with previous designs. The filter array enabled us to subtract dark current drift and attain a clearer light signal. These improvements allowed us to measure bioluminescence reporter-based gene expression in living mammalian cells.

Conclusion: Using our μCIS system for bioluminescence imaging in the future, the device can be implanted in vivo for simultaneous gene expression imaging, behavioral analysis, and optogenetic modulation.

Keywords: bioluminescence; biophotonics; gene expression; image sensors; luciferase; micro-CMOS.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Gene Expression
  • Genes, Reporter
  • Luminescent Measurements*