[Anti-interference hemoglobin analysis system by high performance liquid chromatography]

Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2021 Oct 25;38(5):940-950. doi: 10.7507/1001-5515.202101069.
[Article in Chinese]

Abstract

High performance liquid chromatography (HPLC) is currently the mainstream technology for detecting hemoglobin. Glycated hemoglobin (HbA1c) is a gold indicator for diagnosing diabetes, however, the accuracy of HbA1c test is affected by thalassemia factor hemoglobin F (HbF)/hemoglobin A2 (HbA2) and variant hemoglobin during HPLC analysis. In this study, a new anti-interference hemoglobin analysis system of HPLC is proposed. In this system, the high-pressure three-gradient elution method was improved, and the particle size and sieve plate aperture in the high-pressure chromatography column and the structure of the double-plunger reciprocating series high-pressure pump were optimized. The system could diagnose both HbA1c and thalassemia factor HbF/HbA2 and variant hemoglobin, and the performance of the system was anti-interference and stable. It is expected to achieve industrialization. In this study, the HbA1c and thalassemia factor HbF/HbA2 detection performance was compared between this system and the world's first-line brand products such as Tosoh G8, Bio-Rad Ⅶ and D10 glycosylated hemoglobin analysis system. The results showed that the linear correlation between this system and the world-class system was good. The system is the first domestic hemoglobin analysis system by HPLC for screening of HbA1c and thalassemia factor HbF/HbA2 rapidly and accurately.

高效液相色谱法(HPLC)是目前主流的检测血红蛋白技术,糖化血红蛋白(HbA1c)是诊断糖尿病的金指标,但 HPLC 检测时 HbA1c 易受到地中海贫血因子血红蛋白 F(HbF)/血红蛋白 A2(HbA2)和血红蛋白变异体的干扰而影响检测结果的准确性。本研究基于高效液相色谱法提出一种新型抗干扰血红蛋白分析系统,本系统通过改进高压三梯度洗脱法,优化高压层析柱中微球粒径和筛板孔径,并优化双柱塞往复式串联高压泵的结构,既能检测 HbA1c 又可检测地中海贫血因子 HbF/HbA2 和血红蛋白变异体,抗干扰性好,性能稳定,有望实现产业化。本系统与日本东曹 G8、美国伯乐Ⅶ和 D10 糖化血红蛋白分析系统等世界一线品牌产品进行 HbA1c 和地中海贫血因子 HbF/HbA2 检测性能对比,结果表明,本系统与世界一流系统的线性相关性好,检测性能优异,是国内首次实现快速、精准筛查糖化血红蛋白和地中海贫血因子 HbF/HbA2 的高效液相色谱法血红蛋白分析系统。.

Keywords: glycated hemoglobin; hemoglobin variant; high performance liquid chromatography; thalassemia factor HbF/HbA2.

MeSH terms

  • Chromatography, High Pressure Liquid
  • Fetal Hemoglobin* / analysis
  • Glycated Hemoglobin / analysis
  • Hemoglobin A2* / analysis
  • Hemoglobins

Substances

  • Glycated Hemoglobin A
  • Hemoglobins
  • Hemoglobin A2
  • Fetal Hemoglobin

Grants and funding

工信部 2020 年先进制造业集群高端医疗器械先进制造业集群建设项目(S-2020-C35-500989)