Measurement error due to self-absorption in calibration-free laser-induced breakdown spectroscopy

Anal Chim Acta. 2021 Nov 15:1185:339070. doi: 10.1016/j.aca.2021.339070. Epub 2021 Sep 18.

Abstract

Self-absorption of spectral lines is known to lower the performance of analytical measurements via calibration-free laser-induced breakdown spectroscopy. However, the error growth due to this effect is not clearly assessed. Here we propose a method to quantify the measurement error due to self-absorption based on the calculation of the spectral radiance of a plasma in local thermodynamic equilibrium. Validated through spectroscopic measurements for a binary alloy thin film of compositional gradient, the method evidences that measurement performance lowering due to self-absorption depends on the spectral shape of the analytical transition and on the intensity measurement method. Thus, line-integrated intensity measurements of Stark broadened lines enable accurate analysis, even at large optical thickness, if line width and plasma size are precisely known. The error growth due to self-absorption is significantly larger for line shapes dominated by Doppler broadening and for line-center intensity measurements. The findings present a significant advance in compositional measurements via calibration-free laser-induced breakdown spectroscopy, as they enable straightforward selection of most appropriate analytical lines.

Keywords: Calibration-free; Elemental analysis; Error evaluation; Laser-induced breakdown spectroscopy; Self-absorption.