A computational optimization study of a self-expandable transcatheter aortic valve

Comput Biol Med. 2021 Dec:139:104942. doi: 10.1016/j.compbiomed.2021.104942. Epub 2021 Oct 14.

Abstract

Developing an efficient stent frame for transcatheter aortic valves (TAV) needs thorough investigation in different design and functional aspects. In recent years, most TAV studies have focused on their clinical performance, leaflet design, and durability. Although several optimization studies on peripheral stents exist, the TAV stents have different functional requirements and need to be explicitly studied. The aim of this study is to develop a cost-effective optimization framework to find the optimal TAV stent design made of Ni-Ti alloy. The proposed framework focuses on minimizing the maximum strain occurring in the stent during crimping, making use of a simplified model of the stent to reduce computational cost. The effect of the strut cross-section of the stent, i.e., width and thickness, and the number and geometry of the repeating units of the stent (both influencing the cell size) on the maximum strain is investigated. Three-dimensional simulations of the crimping process are used to verify the validity of the simplified representation of the stent, and the radial force has been calculated for further evaluation. The results suggest the key role of the number of cells (repeating units) and strut width on the maximum strain and, consequently, on the stent design. The difference in terms of the maximum strain between the simplified and the 3D model was less than 5%, confirming the validity of the adopted modeling strategy and the robustness of the framework to improve the TAV stent designs through a simple, cost-effective, and reliable procedure.

Keywords: Finite element analysis; Genetic algorithm; Ni–Ti alloys; Optimization; Shape memory alloy; Transcatheter aortic valve.

MeSH terms

  • Alloys
  • Aortic Valve* / surgery
  • Heart Valve Prosthesis*
  • Prosthesis Design
  • Stents

Substances

  • Alloys