HPLC-potentiometric method for determination of biogenic amines in alcoholic beverages: A reliable approach for food quality control

Food Chem. 2022 Mar 15:372:131288. doi: 10.1016/j.foodchem.2021.131288. Epub 2021 Oct 4.

Abstract

Determination of ten biogenic amines in alcoholic beverages by HPLC coupled to a potentiometric detector for food quality control is herein presented. Biogenic amines were separated by ion-pair chromatography on a C18 column using a gradient mobile phase of acetic acid, acetonitrile, and butane-sulfonic acid. Detection was accomplished by a miniaturized amine-selective electrode. The method was validated following ICH and Eurachem guidelines. Linear regression models provided R2 values from 0.9870 ± 0.0019 to 0.9991 ± 0.0014 for tyramine and cadaverine, respectively. Detection and quantification limits depend on the molecular weight of BAs, ranging from 9.3 to 60.5 and from 31.1 to 202.3 µg L-1 for methylamine and spermine, respectively. Repeatability and intermediate precision showed RSD values lower than 5.8 and 8.3%, respectively. Accuracy of assays yielded recovery values from 86.4 to 109.9%. The biogenic amines content in red wine, white wine, and beer samples were 7.54, 5.24, and 4.58 mg L-1, respectively.

Keywords: Alcoholic beverages; Biogenic amines; Ion-selective electrodes; Liquid chromatography; Potentiometry.

MeSH terms

  • Alcoholic Beverages / analysis
  • Beer / analysis
  • Biogenic Amines* / analysis
  • Chromatography, High Pressure Liquid
  • Quality Control
  • Wine* / analysis

Substances