Retrospective study of deep learning to reduce noise in non-contrast head CT images

Comput Med Imaging Graph. 2021 Dec:94:101996. doi: 10.1016/j.compmedimag.2021.101996. Epub 2021 Sep 30.

Abstract

Purpose: Presented herein is a novel CT denoising method uses a skip residual encoder-decoder framework with group convolutions and a novel loss function to improve the subjective and objective image quality for improved disease detection in patients with acute ischemic stroke (AIS).

Materials and methods: In this retrospective study, confirmed AIS patients with full-dose NCCT head scans were randomly selected from a stroke registry between 2016 and 2020. 325 patients (67 ± 15 years, 176 men) were included. 18 patients each with 4-7 NCCTs performed within 5-day timeframe (83 total scans) were used for model training; 307 patients each with 1-4 NCCTs performed within 5-day timeframe (380 total scans) were used for hold-out testing. In the training group, a mean CT was created from the patient's co-registered scans for each input CT to train a rotation-reflection equivariant U-Net with skip and residual connections, as well as a group convolutional neural network (SRED-GCNN) using a custom loss function to remove image noise. Denoising performance was compared to the standard Block-matching and 3D filtering (BM3D) method and RED-CNN quantitatively and visually. Signal-to-noise ratio (SNR) and contrast-to-noise (CNR) were measured in manually drawn regions-of-interest in grey matter (GM), white matter (WM) and deep grey matter (DG). Visual comparison and impact on spatial resolution were assessed through phantom images.

Results: SRED-GCNN reduced the original CT image noise significantly better than BM3D, with SNR improvements in GM, WM, and DG by 2.47x, 2.83x, and 2.64x respectively and CNR improvements in DG/WM and GM/WM by 2.30x and 2.16x respectively. Compared to the proposed SRED-GCNN, RED-CNN reduces noise effectively though the results are visibly blurred. Scans denoised by the SRED-GCNN are shown to be visually clearer with preserved anatomy.

Conclusion: The proposed SRED-GCNN model significantly reduces image noise and improves signal-to-noise and contrast-to-noise ratios in 380 unseen head NCCT cases.

Keywords: Acute ischemic stroke; CT denoising; Deep learning; Non-contrast head CT.