Percutaneous transaxillary access for endovascular aortic procedures in the multicenter international PAXA registry

J Vasc Surg. 2022 Mar;75(3):868-876.e3. doi: 10.1016/j.jvs.2021.08.089. Epub 2021 Sep 30.

Abstract

Background: The aim of the study was to demonstrate the safety and effectiveness of a suture-mediated vascular closure device to perform hemostasis after an axillary artery access during endovascular procedures on the aortic valve, the aorta and its side branches.

Methods: A physician-initiated, international, multicenter, retrospective registry was designed to evaluate the success rate (VARC-2 reporting standards) of percutaneous transaxillary access closure with a suture-mediated closure device. Secondary end points were minor access vascular complications, transient peripheral nerve injury, stroke, and influence on periprocedural outcomes of puncture technique.

Results: Three hundred thirty-one patients (median age, 76 years; 69.2% males) in 11 centers received a percutaneous transaxillary access during endovascular cardiac (n = 166) or vascular (n = 165) procedures. The closure success rate was 84.6%, with 5 open conversions (1.5%), 45 adjunctive endovascular procedures (13.6%), and 1 nerve injury (0.3%). Secondary closure success was obtained in 325 patients (98%) after 7 bare stenting, 37 covered stenting, and 1 thrombin injection. Introducer sheaths 16F or larger (odds ratio, 3.70; 95% confidence interval, 1.22-11.42) and balloon-assisted hemostasis (odds ratio, 4.45; 95% confidence interval, 1.27-15.68) were associated with closure failure. A threshold of five percutaneous axillary accesses was associated with decreased rates of open conversion, but not with increased primary closure success. Primary closure success was 90.3% in the 175 patients with sheaths smaller than 16F, performed after the first 5 procedures in each center. Temporary nerve injury and stroke were observed in 2% and 4% of patients, respectively.

Conclusions: Percutaneous transaxillary aortic procedures, in selected patients, can be performed with low rates of open conversion. The need for additional endovascular bailout procedures is not negligible when introducers sheaths 16F or larger are required.

Keywords: TAVI; Thoracic aorta aneurysm; Vascular access closure device; Ventricular assist device.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aorta, Thoracic / diagnostic imaging
  • Aorta, Thoracic / surgery*
  • Aortic Diseases / diagnostic imaging
  • Aortic Diseases / surgery*
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / surgery*
  • Axillary Artery*
  • Blood Vessel Prosthesis Implantation* / adverse effects
  • Blood Vessel Prosthesis Implantation* / instrumentation
  • Catheterization, Peripheral* / adverse effects
  • Endovascular Procedures* / adverse effects
  • Endovascular Procedures* / instrumentation
  • Europe
  • Female
  • Heart Valve Diseases / diagnostic imaging
  • Heart Valve Diseases / surgery*
  • Humans
  • Male
  • Postoperative Complications / etiology
  • Punctures
  • Registries
  • Retrospective Studies
  • Risk Assessment
  • Risk Factors
  • Stents
  • Time Factors
  • Treatment Outcome
  • United States