Micro-cavity perfect blackbody composite with good heat transfer towards a flat-plate reference radiation source for thermal imagers

Opt Lett. 2021 Oct 1;46(19):4871-4874. doi: 10.1364/OL.433028.

Abstract

One of the key measures to secure reliable fever screening is to calibrate a thermal imager with an accurate flat-plate blackbody device in real time. We provide durable perfect blackbody plates with both high emissivity of >0.998 and good heat transfer, ideal for a high-precision reference radiation source. Reflectance measurements and heat transfer simulation demonstrate that a micro-cavity composite of a thin resin double layer or resin mixture with thermally conductive filler is an important solution for improving the emissivity and thermal performance of blackbody plates.