Calibration Technique for Suppressing Residual Etalon Artifacts in Slit-Averaged Raman Spectroscopy

Appl Spectrosc. 2022 Feb;76(2):255-261. doi: 10.1177/00037028211046643. Epub 2021 Oct 1.

Abstract

Back-illuminated charged-coupled device (BI-CCD) arrays increase quantum efficiency but also amplify etaloning, a multiplicative, wavelength-dependent fixed-pattern effect. When spectral data from hundreds of BI-CCD rows are combined, the averaged spectrum will generally appear etalon-free. This can mask substantial etaloning at the row level, even if the BI-CCD has been treated to suppress the effect. This paper compares two methods of etalon correction, one with simple averaging and one with row-by-row calibration using a fluorescence standard. Two BI-CCD arrays, both roughened by the supplier to reduce etaloning, were used to acquire Raman spectra of murine bone specimens. For one array, etaloning was the dominant source of noise under the exposure conditions chosen, even for the averaged spectrum across all rows; near-infrared-excited Raman peaks were noticeably affected. In this case, row-by-row calibration improved the spectral quality of the average spectrum. The other CCD's performance was shot-noise limited and therefore received no benefit from the extra calibration. The different results highlight the importance of checking for and correcting row-level fixed pattern when measuring weak Raman signals in the presence of a large fluorescence background.

Keywords: Fixed pattern; Raman spectroscopy; calibration; etalon effect.

MeSH terms

  • Animals
  • Artifacts*
  • Calibration
  • Mice
  • Spectrum Analysis, Raman*