Validation of Numerically Predicted Shear Stress-dependent Dissipative Losses Within a Rotary Blood Pump

ASAIO J. 2021 Oct 1;67(10):1148-1158. doi: 10.1097/MAT.0000000000001488.

Abstract

Computational fluid dynamics find widespread application in the development of rotary blood pumps (RBPs). Yet, corresponding simulations rely on shear stress computations that are afflicted with limited resolution while lacking validation. This study aimed at the experimental validation of integral hydraulic properties to analyze global shear stress resolution across the operational range of a novel RBP. Pressure head and impeller torque were numerically predicted based on Unsteady Reynolds-averaged Navier-Stokes (URANS) simulations and validated on a testbench with integrated sensor modalities (flow, pressure, and torque). Validation was performed by linear regression and Bland-Altman analysis across nine operating conditions. In power loss analysis (PLA), in silico hydraulic power losses were derived based on the validated hydraulic quantities and balanced with in silico shear-dependent dissipative power losses. Discrepancies among both terms provided a measure of in silico shear stress resolution. In silico and in vitro data correlated with low discordance in pressure (r = 0.992, RMSE = 1.02 mmHg), torque (r = 0.999, RMSE = 0.034 mNm), and hydraulic power losses (r = 0.990, RMSE = 0.015W). PLA revealed numerically predicted dissipative losses to be up to 34.4% smaller than validated computations of hydraulic losses. This study confirmed the suitability of URANS settings to predict integral hydraulic properties. However, numerical credibility was hampered by lacking resolution of shear-dependent dissipative losses.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Computer Simulation
  • Heart-Assist Devices*
  • Hydrodynamics
  • Models, Cardiovascular
  • Stress, Mechanical