This work shows the comparison between Geant4 low energy electromagnetic physics lists G4EmLi-vermorePhysics, G4EmPenelopePhysics, G4EmLowEPPhysics, and G4EmDNAPhysics_option2 when simulating the energy deposition of low mono-energetic electrons and β- emitted from 90Y isotope. The simulation time and influence of production cut were considered. In the sense of balance between the accuracy and computer resource, G4EmPenelopePhysics can be proposed as the best physics model for our future Treatment Planning System (TPS) for treating liver cancer using 90Y microsphere radioembolization therapy.
Keywords: Geant4; Low energy electromagnetic models; Yttrium-90 radioembolization.
Copyright © 2021 Elsevier Ltd. All rights reserved.