Spurious phase correction in rapid metabolic imaging

J Magn Reson. 2021 Nov:332:107065. doi: 10.1016/j.jmr.2021.107065. Epub 2021 Sep 8.

Abstract

IDEAL-type magnetic resonance spectroscopic imaging (MRSI) sequences require the acquisition of several datasets using optimized sampling in the time domain to reconstruct metabolite maps. Each unitary scan consists of a selective slice (2D) or slab (3D) excitation followed by an evolution time and then the acquisition of the spatially encoded signal. It is critical that the phase variation during the evolution time for each scan is only dependent on chemical shifts. In this paper, we described the apparition of spurious phase due to either the transmit or the receive frequency. The presence of this unwanted phase depends on (i) where the commutation between these two frequencies is performed and (ii) how it is done, as there are two phase commutation modes: continuous and coherent. We present the correction needed in function of the different cases. It appears that some solutions are universal. However, it is critical to know which case is implemented on the MRI scanner, which is not always easy information to have. We illustrated several cases with our preclinical MRI by using the IDEAL spiral method on a 13C phantom.

Keywords: Frequency commutation; MRSI; Multi-shot imaging;; Phase correction; Phase-coherent; Phase-continuous.

Publication types

  • Clinical Conference
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Brain*
  • Imaging, Three-Dimensional
  • Magnetic Resonance Imaging
  • Phantoms, Imaging
  • Phase Variation*