Effect of Stem Positioning on Biomechanical Performance of a Novel Cementless Short-Stem Canine Total Hip Implant

Vet Comp Orthop Traumatol. 2022 Jan;35(1):1-9. doi: 10.1055/s-0041-1735290. Epub 2021 Sep 21.

Abstract

Objective: The aim of this study was to evaluate the effect of stem positioning on the biomechanical performance of a novel, collared, short-stem total hip implant under compression and torsion ex vivo.

Study design: Six canine cadaveric femurs were implanted with a collared short-stem femoral implant. Canal flare index (CFI), stem angle, absolute and relative cut heights and relative size were measured radiographically and used as independent variables. Biomechanical performance of the construct was evaluated using physiologic loading (loading) and supraphysiologic loading (failure) protocols.

Results: During loading protocols, compressive stiffness was influenced by absolute cut height (p = 0.018). During failure protocols, peak torque was influenced by CFI (p = 0.004) and craniocaudal relative size (p = 0.005). Peak load and torsional stiffness were not impacted by any of the radiographic variables (p > 0.05). Three of six femurs developed longitudinal fractures originating at the medial calcar at the time of failure.

Conclusion: The biomechanical performance of the collared short-stem implant was positively impacted by preserving more of the femoral neck, having a higher CFI and using a smaller implant size relative to the femoral neck isthmus.

MeSH terms

  • Animals
  • Arthroplasty, Replacement, Hip* / veterinary
  • Biomechanical Phenomena
  • Dogs
  • Femur / diagnostic imaging
  • Femur / surgery
  • Femur Neck
  • Hip Prosthesis* / veterinary
  • Prosthesis Design / veterinary
  • Torque