Tunable and Contamination-Free Injection with Microfluidics by Stepinjection

Anal Chem. 2021 Oct 5;93(39):13112-13117. doi: 10.1021/acs.analchem.1c02721. Epub 2021 Sep 21.

Abstract

Droplet microfluidics with picoinjection provides significant advantages to multistep reactions and screenings. The T-junction design for picoinjection is convenient in adding picoliter reagents into passing droplets to initiate reactions. However, conventional picoinjectors face difficulties in eliminating cross-contamination between droplets, preventing them from widespread use in sensitive biological and molecular assays. Here, we introduce stepinjection, which uses a T-junction with a stepped channel design to elevate the diffusional buffer zone into the main channel and consequently increases the pressure difference between droplets and the inlet of the injection channel. To demonstrate the stepinjector's ability to perform contamination-sensitive enzymatic assays, we inject casein fluorescein isothiocyanate (FITC-casein) into a mixture of savinase and savinase-free (labeled with a red fluorescent dye) droplets. We observe no cross-contamination using stepinjection but find a severe cross-talk using an optimal picoinjection design. We envision that the simple, tunable, and reliable stepinjector can be easily integrated in various droplet processing devices, and facilitate various biomedical and biochemical applications including multiplex digital PCR, single-cell sequencing, and enzymatic screening.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Equipment Contamination*
  • Microfluidic Analytical Techniques
  • Microfluidics*