A taper-fit junction to improve long bone reconstruction: A parametric In Silico model

J Mech Behav Biomed Mater. 2021 Dec:124:104790. doi: 10.1016/j.jmbbm.2021.104790. Epub 2021 Aug 26.

Abstract

Purpose: Critical size long bone defects represent a clinical challenge in orthopaedic surgery. Various grafting techniques have been developed through the years, but they all present several downsides. A key requirement of all grafting techniques is the achievement of a continuous interface between host bone and graft to enhance both biological processes and mechanical stability. This study used a parametric in silico model to quantify the biomechanical effect of the inaccuracies inherent to current osteotomy techniques, and to test a new concept of accurate taper-fit junction that may improve the biomechanical parameters of the reconstruction under load.

Methods: A population-based in-silico 3D model of the reconstruction of a long bone defect was built to represent a defect of the femoral mid-diaphysis. To fix the reconstruction a titanium plate was placed on the lateral aspect of the reconstruction. The model was modified to (i) quantify the biomechanical consequences of actual inaccuracies in the realization of a flat host-graft interface, (ii) compare the contact behaviour and bone strains among different taper angles of the new design and the current host-graft flat interface, (iii) evaluate the robustness of the taper-fit design to inter-subject variability in bone geometry and defect length.

Results: The influence of 2° single-plane misalignments of the host-graft interface is highly dependent on the misalignment orientation with respect to the metal plate. For some misalignment orientations, tangential micromotions of contact interfaces exceeded alert thresholds. When the angle of the taper-fit host-graft junction is changed from 10° to 30° and the results obtained are compared with the planar case, the overall stiffness is almost preserved, the bone strains are almost unchanged with safety factors higher than five, and full contact closure around the host-graft junction is achieved at 20°. Similarly, contact pressures decrease almost linearly with a 20% decrease at 30°. The host-graft micro motions are almost unchanged in both value and distribution up to 20° and never exceed the warning threshold of 50 μm.

Conclusions: The present in silico study developed quantitative biomechanical evidence that an osteotomy performed with attention to the perpendicularity of the cut planes is needed to reduce the risk of mismatch and possible complications of long bone reconstructions, and that a new concept of a taper-fit junction may improve the biomechanical environment of the interface between the graft and the host bone. The optimal taper-fit configuration is suggested to be around a 20° taper angle. These results will serve as an input to conduct exvivo experiments to further corroborate the proposed taper-fit junction concept and to refine its surgical implementation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biomechanical Phenomena
  • Bone Plates
  • Computer Simulation
  • Femur* / surgery
  • Humans
  • Osteotomy
  • Plastic Surgery Procedures*