Comparison Between Conventional Flash and Off-Response Intraoperative Visual Evoked Potential Monitoring for Endoscopic Endonasal Surgery

Oper Neurosurg (Hagerstown). 2021 Nov 15;21(6):516-522. doi: 10.1093/ons/opab329.

Abstract

Background: Intraoperative flash stimulation visual evoked potential (VEP) monitoring has been used for endoscopic endonasal approach (EEA). Recently, off-response VEP, which is recorded when the light stimulus is turned off, was introduced to monitor visual function intraoperatively.

Objective: To evaluate off-response VEP monitoring in comparison with the conventional flash stimulation VEP monitoring for EEA.

Methods: From March 2015 to March 2020, 70 EEA surgeries with intraoperative VEP monitoring (140 eyes) were performed. Light stimuli were delivered by a pair of goggle electrodes. Recording electrodes were placed on the scalp over the occipital region. The warning signal was prompted by a reduction of the peak-to-peak amplitude of the VEP by more than 50% compared to the initial amplitude. Visual function was assessed pre- and postoperatively. Results of flash and off-response VEP monitoring were compared.

Results: VEP was recorded in 134 eyes. Warning signal occurred in 23 eyes (transient in 17 eyes and permanent in 6 eyes). Two eyes showed permanent VEP attenuation for flash VEP monitoring, in which one patient had postoperative visual function deterioration. Four eyes showed permanent VEP attenuation for off-response VEP monitoring, where 2 patients had postoperative visual function deterioration. Sensitivity, specificity, positive predictive value, and negative predictive value were 100%, 98.9%, 50%, and 100%, respectively, for flash stimulation VEP, and 100%, 97.8%, 50%, and 100%, respectively, for off-response VEP.

Conclusion: VEP monitoring was useful to monitor visual function in EEA surgery. Off-response VEP monitoring was not inferior to conventional flash stimulation VEP monitoring.

Keywords: Endoscopic endonasal approach; Flash stimulation; Intraoperative neurophysiological monitoring; Off-response; Visual evoked potential.