Mechanical circulatory support devices (MCSDs), although proved to be a pillar in the clinical setting of advanced heart failure, are afflicted by thromboembolic complications. Shear-mediated platelet activation has been recognized to drive thromboembolic events in patients implanted with MCSDs. Despite this, to date, a clinically reliable diagnostic test for assessing platelet response to stress stimuli is still missing. Here, we describe and apply the previously developed device thrombogenicity emulation methodology to the design of a microfluidic platform able to replicate shear stress profiles representative of MCSDs. The device-specific shear-mediated platelet activation is finally assessed by the platelet activity state assay, which measures real-time thrombin production, as a marker of platelet activation level. This technique can be employed to emulate the shear stress patterns of different MCSDs, such as mechanical heart valves, ventricular assist devices, and stents.
Keywords: Mechanical circulatory support devices; Microfluidics; Platelet activation; Shear stress patterns; Thrombogenicity.
© 2022. Springer Science+Business Media, LLC, part of Springer Nature.