A rare complication of pelvic perforation by an excessive medial slide of the helical blade after treatment of an intertrochanteric fracture with proximal femoral nail anti-rotation: A case report and literature review

Chin J Traumatol. 2022 Mar;25(2):118-121. doi: 10.1016/j.cjtee.2021.08.007. Epub 2021 Aug 12.

Abstract

Intertrochanteric fractures have become a severe public health problem in elderly patients. Proximal femoral nail anti-rotation (PFNA) is a commonly used intramedullary fixation device for unstable intertrochanteric fractures. Pelvic perforation by cephalic screw is a rare complication. We reported an 84-year-old female who fell at home and sustained an intertrochanteric fracture. The patient underwent surgery with PFNA as the intramedullary fixation device. Routine postoperative examination revealed medial migration of the helical blade that eventually caused pelvic perforation. We performed a cemented total hip arthroplasty as the savage procedure. At the latest follow-up of 12 months after total hip arthroplasty, the patient had no pain or loosening of the prosthesis in the left hip. Pelvic perforation should be considered when choosing PFNA as the intramedullary fixation device, especially in patients with severe osteoporosis wherein the helical blade can be easily inserted during the operation. The lack of devices to avoid oversliding of the helical blade in PFNA is an unreported cause of this complication and should be considered in such cases.

Keywords: Case report; Complication; Hip fractures; Intertrochanteric fractures; Pelvic perforation; Proximal femoral nail anti-rotation.

Publication types

  • Case Reports
  • Review

MeSH terms

  • Aged
  • Aged, 80 and over
  • Bone Nails / adverse effects
  • Female
  • Femur
  • Fracture Fixation, Intramedullary* / adverse effects
  • Hip Fractures* / surgery
  • Humans
  • Retrospective Studies
  • Treatment Outcome