Proof-of-Concept of a Sensor-Based Evaluation Method for Better Sensitivity of Upper-Extremity Motor Function Assessment

Sensors (Basel). 2021 Sep 3;21(17):5926. doi: 10.3390/s21175926.

Abstract

In rehabilitation, the Fugl-Meyer assessment (FMA) is a typical clinical instrument to assess upper-extremity motor function of stroke patients, but it cannot measure fine changes of motor function (both in recovery and deterioration) due to its limited sensitivity. This paper introduces a sensor-based automated FMA system that addresses this limitation with a continuous rating algorithm. The system consists of a depth sensor (Kinect V2) and an algorithm to rate the continuous FM scale based on fuzzy inference. Using a binary logic based classification method developed from a linguistic scoring guideline of FMA, we designed fuzzy input/output variables, fuzzy rules, membership functions, and a defuzzification method for several representative FMA tests. A pilot trial with nine stroke patients was performed to test the feasibility of the proposed approach. The continuous FM scale from the proposed algorithm exhibited a high correlation with the clinician rated scores and the results showed the possibility of more sensitive upper-extremity motor function assessment.

Keywords: depth sensor; fuzzy inference system; rehabilitation; sensitivity; sensor-based motor function assessment.

MeSH terms

  • Algorithms
  • Humans
  • Recovery of Function
  • Stroke Rehabilitation*
  • Stroke* / diagnosis
  • Upper Extremity