A Neural Recording and Stimulation Chip with Artifact Suppression for Biomedical Devices

J Healthc Eng. 2021 Aug 27:2021:4153155. doi: 10.1155/2021/4153155. eCollection 2021.

Abstract

This paper presents chip implementation of the integrated neural recording and stimulation system with stimulation-induced artifact suppression. The implemented chip consists of low-power neural recording circuits, stimulation circuits, and action potential detection circuits. These circuits constitute a closed-loop simultaneous neural recording and stimulation system for biomedical devices, and a proposed artifact suppression technique is used in the system. Moreover, this paper also presents the measurement and experiment results of the implemented 4-to-4 channel neural recording and stimulation chip with 0.18 µm CMOS technology. The function and efficacy of simultaneous neural recording and stimulation is validated in both in vivo and animal experiments.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Action Potentials
  • Animals
  • Artifacts*