The combined use of thoracic and abdominal aortic stent grafts for endovascular repair of wide neck abdominal aortic aneurysms in high risk patients

Ann Vasc Surg. 2021 Nov:77:350.e1-350.e7. doi: 10.1016/j.avsg.2021.05.038. Epub 2021 Aug 23.

Abstract

Introduction: Treatment of abdominal aortic aneurysms (AAA) with large (28 mm to 34 mm) and wide diameter (> 35 mm) necks remains a challenge in patients who are high-risk candidates for open repair. While several case reports describe the use of a thoracic stent graft in conjunction with a traditional modular bifurcated stent graft, most patients do not have the aortic length to accommodate such a configuration. We present our experience utilizing a distal unibody bifurcated aortic stent graft (Endologix, Irvine, CA) in conjunction with a proximal thoracic aortic stent graft (Medtronic, Minneapolis, MN) to treat wide-necked non-ruptured AAAs in patients who were otherwise poor candidates for open or fenestrated repair.

Methods: A single center retrospective review of patients treated with a combination of a distal unibody bifurcated aortic stent graft and a proximal thoracic aortic stent graft extension from 2013 to 2019 was performed. Demographics, perioperative details and long-term outcomes were collected and summarized. Standard statistical methods were utilized.

Results: We identified 7 patients who underwent this procedure during the study interval. Of these, all 7 (100%) were male with an average age of 69.1 ± 5.1 years. Average Charlson Comorbidity Index was 5.0. Average pre-operative maximum aortic and neck diameters were 57.9 mm (± 5.8) and 37.4 mm (± 4.5) respectively. All patients underwent repair with a distal 28 mm diameter unibody bifurcated aortic stent graft and proximal extension with a thoracic aortic stent graft that ranged from 40 to 46 mm in diameter. Technical success was achieved in all 7 patients. There were no perioperative mortalities or aorta-related deaths. Follow up was a mean of 1.98 years with a mean survival of 4.75 years (± 0.86). One patient required an aneurysm-related intervention for a late type III endoleak.

Conclusion: The combined use of thoracic and abdominal aortic stent grafts is a safe and effective endovascular method to treat high-risk surgical candidates with wide-necked AAAs.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Aortic Aneurysm, Abdominal / diagnostic imaging
  • Aortic Aneurysm, Abdominal / surgery*
  • Blood Vessel Prosthesis Implantation / adverse effects
  • Blood Vessel Prosthesis Implantation / instrumentation*
  • Blood Vessel Prosthesis*
  • Databases, Factual
  • Endoleak / etiology
  • Endoleak / therapy
  • Endovascular Procedures / adverse effects
  • Endovascular Procedures / instrumentation*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Prosthesis Design
  • Retreatment
  • Retrospective Studies
  • Risk Assessment
  • Risk Factors
  • Stents*
  • Treatment Outcome