Bearing fixing: A new computer algorithm method for subjective determination of astigmatism

Ophthalmic Physiol Opt. 2021 Sep;41(5):1060-1068. doi: 10.1111/opo.12848. Epub 2021 Aug 13.

Abstract

Purpose: This article aims to illustrate the principles of bearing fixing, a new computer-controlled procedure for subjective determination of astigmatism, and to compare it with conventional clinician-controlled Jackson crossed-cylinder refraction.

Methods: The principles and method for bearing fixing are described. Astigmatism was measured using bearing fixing and Jackson cross-cylinder twice on 20 visually normal participants aged between 18 and 34 years. After final sphere adjustment, acuity measurements were made with each refraction estimate.

Results: Bearing fixing results could be obtained for all participants. Mean bearing fixing cylinder magnitude was slightly higher than Jackson cross-cylinder measures of cylinder magnitude, by 0.05D. Using vector analysis to take into account cylinder power and axis, mean bearing fixing astigmatism was not significantly different from cross-cylinder astigmatism, but did have higher test-retest variability (p < 0.05). Acuity with bearing fixing and cross-cylinder corrections did not significantly differ in average value or repeatability.

Conclusions: Bearing fixing is a feasible method for subjective determination of astigmatism, and there are a number of potential improvements to the method which could make it faster and more precise.

Keywords: astigmatism; computer controlled refraction; optics; refraction algorithm.

MeSH terms

  • Adolescent
  • Adult
  • Algorithms
  • Astigmatism* / diagnosis
  • Computers
  • Humans
  • Refraction, Ocular
  • Visual Acuity