Factors affecting signal quality in implantable cardiac monitors with long sensing vector

J Arrhythm. 2021 Jun 21;37(4):1061-1068. doi: 10.1002/joa3.12585. eCollection 2021 Aug.

Abstract

Purpose: Electrical artefacts are frequent in implantable cardiac monitors (ICMs). We analyzed the subcutaneous electrogram (sECG) provided by an ICM with a long sensing vector and factors potentially affecting its quality.

Methods: Consecutive ICM recipients underwent a follow-up where demographics, body mass index (BMI), implant location, and surface ECG were collected. The sECG was then analyzed in terms of R-wave amplitude and P-wave visibility.

Results: A total of 84 patients (43% female, median age 68 [58-76] years) were enrolled at 3 sites. ICMs were positioned with intermediate inclination (n = 44, 52%), parallel (n = 35, 43%), or perpendicular (n = 5, 6%) to the sternum. The median R-wave amplitude was 1.10 (0.72-1.48) mV with P waves readily visible in 69.2% (95% confidence interval, CI: 57.8%-79.2%), partially visible in 23.1% [95% CI: 14.3%-34.0%], and never visible in 7.7% [95% CI: 2.9%-16.0%] of patients. Men had higher R-wave amplitudes compared to women (1.40 [0.96-1.80] mV vs 1.00 [0.60-1.20] mV, P = .001), while obese people tended to have lower values (0.80 [0.62-1.28] mV vs 1.10 [0.90-1.50] mV, P = .074). The P-wave visibility reached 86.2% [95% CI: 68.3%-96.1%] in patients with high-voltage P waves (≥0.2 mV) at surface ECG. The sECG quality was not affected by implant site.

Conclusion: In ordinary clinical practice, ICMs with long sensing vector provided median R-wave amplitude above 1 mV and reliable P-wave visibility of nearly 70%, regardless of the position of the device. Women and obese patients showed lower but still very good signal quality.

Keywords: P‐wave visibility; R‐wave amplitude; implantable cardiac monitor; implantable loop recorder; long sensing vector.