Comprehensive Analytical Modelling of an Absolute pH Sensor

Sensors (Basel). 2021 Jul 30;21(15):5190. doi: 10.3390/s21155190.

Abstract

In this work, we present a comprehensive analytical model and results for an absolute pH sensor. Our work aims to address critical scientific issues such as: (1) the impact of the oxide degradation (sensing interface deterioration) on the sensor's performance and (2) how to achieve a measurement of the absolute ion activity. The methods described here are based on analytical equations which we have derived and implemented in MATLAB code to execute the numerical experiments. The main results of our work show that the depletion width of the sensors is strongly influenced by the pH and the variations of the same depletion width as a function of the pH is significantly smaller for hafnium dioxide in comparison to silicon dioxide. We propose a method to determine the absolute pH using a dual capacitance system, which can be mapped to unequivocally determine the acidity. We compare the impact of degradation in two materials: SiO2 and HfO2, and we illustrate the acidity determination with the functioning of a dual device with SiO2.

Keywords: analytical model; depletion width; nano-biosensor; oxide degradation; pH sensor modelling and simulations.

MeSH terms

  • Biosensing Techniques*
  • Electric Capacitance
  • Hydrogen-Ion Concentration
  • Oxides
  • Silicon Dioxide*

Substances

  • Oxides
  • Silicon Dioxide