Agreement between optical coherence and Scheimpflug tomography: Vault measurements and reproducibility after implantable collamer lens implantation

J Fr Ophtalmol. 2021 Nov;44(9):1370-1380. doi: 10.1016/j.jfo.2021.03.007. Epub 2021 Jul 28.

Abstract

Purpose: To evaluate the agreement between Scheimpflug tomography (Pentacam, Oculus) and anterior segment optical coherence tomography (AS-OCT, RTVue 100, Optovue) as well as the reproducibility of each technique in assessing the lens-ICL distance (vault) after implantable collamer lens (ICL) implantation.

Methods: The vault was measured manually with Scheimpflug tomography and AS-OCT. Intraclass correlation coefficients (ICC) and Bland-Altman plots were used to determine the reproducibility of measurements and the agreement between them. Multivariate regression analysis was performed to identify predictors of differences in vault measurements between devices.

Results: 80 eyes of 46 ICL patients were analyzed. Mean patient age was 33.8±7.4 years (range, 21 to 51), and 27 (59%) were women. The preoperative spherical equivalent refraction ranged from -26 to 8.5 diopters (D). The mean vault measured by AS-OCT was 558.8±240.2μm (range, 162 - 1220) vs. 430.1±215.3μm (range, 0 to 1070) by Pentacam. AS-OCT yielded vault values, on average, 128.1±64.6μm higher than the Pentacam (range, -22 to 293μm). The ICC values for consistency and absolute agreement were 0.960 and 0.928, respectively. Reproducibility of vault measurements was excellent for both devices (ICC≥0.946). Four parameters-vault, pupil diameter (PD), PD differences between devices, and preoperative spherical equivalent refraction-were significant predictors of differences in vault measurements (adjusted-R2=0.412; P<0.001).

Conclusions: Agreement between AS-OCT and Pentacam for vault measurement was good. Vaults were higher when measured by AS-OCT. These differences are clinically relevant, and therefore these devices cannot be used interchangeably to measure vault.

Keywords: AS-OCT; Agreement; Concordance; Implantable collamer lens; Pentacam; Reproducibility; Reproductibilité; Vault.