Fatigue performance of prosthetic screws used in dental implant restorations: Rolled versus cut threads

J Prosthet Dent. 2021 Sep;126(3):406.e1-406.e8. doi: 10.1016/j.prosdent.2021.06.035. Epub 2021 Jul 24.

Abstract

Statement of problem: Cold rolling is widely used for screw thread manufacturing in industry but is less common in implant dentistry, where cutting is the preferred manufacturing method.

Purpose: The purpose of this in vitro study was to compare the surface finish and mechanical performance of a specific model of prosthetic screw used for direct restorations manufactured by thread rolling and cutting.

Material and methods: The thread profiles were measured in an optical measuring machine, the residual stresses in an X-ray diffractometer, the surface finish in a scanning electron microscope, and then fatigue and static load tests were carried out in a direct stress test bench according to the International Organization for Standardization (ISO) 14801. Finally, linear regression models and 95% interval confidence bands were calculated and compared through ANCOVA for fatigue tests while the t test was used for statistical comparisons (α=.05).

Results: The surface finish was smoother, and compressive residual stresses were higher for the roll-threaded screws. Linear regression models showed a fatigue life 9 times higher for roll-threaded screws (P=1) without affecting static behavior, which showed statistically similar static strengths (P=.54). However, the thread profile in the roll-threaded screws was not accurately reproduced, but this should be easily corrected in future prototypes.

Conclusions: Rolling was demonstrated to be a better thread-manufacturing process for prosthetic screws, producing improved surface quality and fatigue behavior.