Modified Reduced Field-of-View Diffusion-Weighted Magnetic Resonance Imaging of the Prostate: Comparison With Reduced Field-of-View Imaging and Single Shot Echo-Planar Imaging

J Comput Assist Tomogr. 2021 May-Jun;45(3):367-373. doi: 10.1097/RCT.0000000000001156.

Abstract

Objective: The objective of this study was to compare the image quality and apparent diffusion coefficient (ADC) of diffusion-weighted imaging (DWI) with modified reduced field of view (FOV) based on 2-dimensional (2D)-selective radiofrequency excitations by tilting the excitation plane in prostate with reduced FOV using parallel-transmit-accelerated 2D-selective radiofrequency excitation and single-shot echo planar imaging (ssEPI).

Methods: Fifty patients who underwent multiparametric magnetic resonance imaging including 3 DWIs were included. Two observers independently performed qualitative image analyses using 5-point scale. Apparent diffusion coefficient measurements were performed for quantitative analysis.

Results: Modified reduced FOV provided the highest qualitative scores for all categories compared with reduced FOV and ssEPI (P < 0.000). Both reduced FOV DWIs showed higher ADC values compared with ssEPI (P < 0.001); however, the ADC ratios between the lesion and peripheral zone were not significantly different (all P > 0.05).

Conclusions: The modified reduced FOV DWI showed better overall image quality, differentiability of anatomic regions, and lesion conspicuity with fewer artifacts compared with DWI with reduced FOV and ssEPI.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Diffusion Magnetic Resonance Imaging / methods*
  • Echo-Planar Imaging
  • Humans
  • Male
  • Middle Aged
  • Multiparametric Magnetic Resonance Imaging
  • Observer Variation
  • Prostate / diagnostic imaging*
  • Radiographic Image Interpretation, Computer-Assisted / methods*
  • Retrospective Studies
  • Signal-To-Noise Ratio