FDG-CT/PET false positive case in hip prosthesis: a clue to avoid error

Radiol Case Rep. 2021 Jul 10;16(9):2601-2604. doi: 10.1016/j.radcr.2021.06.039. eCollection 2021 Sep.

Abstract

A 62 years old woman 6 months after left total hip prosthesis referred to our institution for persistent pain and warm, stiff, and swollen joint. 18F-FDG CT/PET Images showed an intense focal uptake corresponding to the external margin of inter-trochanteric region of prosthesis and inside the stem inferiorly, but common decision was to reconstruct PET images without attenuation correction and now showed a complete and unexpected disappearance of focal and pathological FDG uptake. This case shows the potential propagation of CT artifacts into PET emission data close to metal implants and should be taken in account together to SUV values.

Keywords: Artifacts; Attenuation correction; FDG-PET; Joint replacement.