Nonlinear reconstruction of coded spectral X-ray CT based on material decomposition

Opt Express. 2021 Jun 21;29(13):19319-19339. doi: 10.1364/OE.426732.

Abstract

Coded spectral X-ray computed tomography (CT) based on K-edge filtered illumination is a cost-effective approach to acquire both 3-dimensional structure of objects and their material composition. This approach allows sets of incomplete rays from sparse views or sparse rays with both spatial and spectral encoding to effectively reduce the inspection duration or radiation dose, which is of significance in biological imaging and medical diagnostics. However, reconstruction of spectral CT images from compressed measurements is a nonlinear and ill-posed problem. This paper proposes a material-decomposition-based approach to directly solve the reconstruction problem, without estimating the energy-binned sinograms. This approach assumes that the linear attenuation coefficient map of objects can be decomposed into a few basis materials that are separable in the spectral and space domains. The nonlinear problem is then converted to the reconstruction of the mass density maps of the basis materials. The dimensionality of the optimization variables is thus effectively reduced to overcome the ill-posedness. An alternating minimization scheme is used to solve the reconstruction with regularizations of weighted nuclear norm and total variation. Compared to the state-of-the-art reconstruction method for coded spectral CT, the proposed method can significantly improve the reconstruction quality. It is also capable of reconstructing the spectral CT images at two additional energy bins from the same set of measurements, thus providing more spectral information of the object.