High-power near-concentric Fabry-Perot cavity for phase contrast electron microscopy

Rev Sci Instrum. 2021 May 1;92(5):053005. doi: 10.1063/5.0045496.

Abstract

Transmission electron microscopy (TEM) of vitrified biological macromolecules (cryo-EM) is limited by the weak phase contrast signal that is available from such samples. Using a phase plate would thus substantially improve the signal-to-noise ratio. We have previously demonstrated the use of a high-power Fabry-Perot cavity as a phase plate for TEM. We now report improvements to our laser cavity that allow us to achieve record continuous wave intensities of over 450 GW/cm2, sufficient to produce the optimal 90° phase shift for 300 keV electrons. In addition, we have performed the first cryo-EM reconstruction using a laser phase plate, demonstrating that the stability of this laser phase plate is sufficient for use during standard cryo-EM data collection.